The launch cord used for the field calibration must not be disconnected from the source otherwise the
calibration is invalid.
Fig. 9 Test Setup According to Reference Method with One Test Cord
4.3 Set Autotest Pref
Autotest is the test mode used most often. The Autotest allows various presettings:
Automatic saving
Passed/Failed Rating
1. On start screen of Display handset (DH) select option Preferences.
2. Select option Autotest Pref.
3. Now enter the desired preferences in the opening screen.
NOTE
Please note that Autotest Pref entered by you must be in compliance with the requirements of the
certification to be carried out.
The certification of fiber cabling is usually performed according to TIA-568 or ISO 11801 standards; in
doing so, the passed/failed criterion is based on cable length, number of splices and connections in the
fiber line.
In addition to these criteria, the FiberTEK III can also be configured for application-specific fiber tests.
These fiber tests are particular beneficial when determining whether a fiber line supports certain types of
network technology, such as 100BaseSX or 1000BaseLX.
4.3.1 Selection of Fiber Test Standard
1. Use Arrow Keys to navigate on ready screen to "LASER" display and confirm with Enter.
2. Use Arrow Keys to select the desired fiber typeset standard and confirm with Enter.
LS
PM
Launch cord
LS
PM
Launch cord
Link under test
Measured Loss