Thermometer User Manual
Page 21 
ACV High Frequency Calibration 
Calibrating the ACV Frequency Response 
1. 400V Range 
 Set the calibrator to generate 200.00V / 10KHz ±0.07% at least, adjust VC3 to 
 have a reading of 200.50± 20 count . 
2. 40V Range 
 Set the calibrator to generate 20.000V /10KHz±0.07% at least, adjust VC2 to 
 have a reading of 20.050± 20 count . 
3. 4V Range 
 Set the calibrator to generate 2.0000V /10KHz±0.07% at least, adjust VC1 to 
 have a reading of 2.0050± 20 count . 
Form Number TM61633-5  Rev 5 Dec 2006 