Thermometer User Manual
Page 17 
Table 9. AC mA Test 
Input Readings 
Step Current Frequency  61-633  61-635 
1  36.000mA  50 Hz  35.488 to 36.512  35.632 to 36.368 
2  36.000mA  400 Hz  35.488 to 36.512  35.632 to 36.368 
3  360.00mA  50 Hz  354.88 to 365.12  356.32 to 363.68 
4  360.00mA  400Hzz  354.88 to 365.12  356.32 to 363.68 
Testing the Amp (A) Function 
To verify the accuracy in the ampere (A) measurement function, do the 
Following: 
1. Connect the Calibrator to the A and COM inputs of the Meter. 
2. Turn the rotary switch to A 
3. Apply the inputs for steps 1-2 in Table 10. 
4. For each input, compare the readings on the Meter display to the display readings in 
 Table 10. 
5. If the display reading falls outside of the range shown in Table 10, the meter does not 
 meet specification. 
Table 10. DCA Test 
Input Readings 
Step Current 61-633  61-635 
1  3.6000A  3.5780 to 3.6220  3.5888 to 3.6112 
2  9.000A  8.9150 to 9.0850  8.9420 to 9.0580 
6. Press the blue button on the Meter to toggle to the AC amp measurement function. 
7. Apply the inputs for steps 1-4 in Table 11. 
8. For each input, compare the reading on the Meter display to the display readings in 
 Table 11. 
9. If the display reading falls outside of the range shown in Table 11, the meter does not 
 meet specification. 
Table 11. ACA Test 
Input Readings 
Step Current Frequency  61-633  61-635 
1  3.6000A  50 Hz  3.5488 to 3.6512  3.5632 to 3.6368 
2  3.6000A  400 Hz  3.5488 to 3.6512  3.5632 to 3.6368 
3  9.000A  50 Hz  8.8120 to 9.1888  8.8480 to 9.1520 
4  9.000A  400 Hz  8.8120 to 9.1888  8.8480 to 9.1520 
Testing the Frequency Function 
To verify the accuracy of the Meter’s frequency function, do the following: 
1. Connect the Calibrator to the VΩ and COM inputs on the Meter. 
Note: The accuracy of the Calibrator’s frequency function must be appropriate for the specified accuracy of the 
Meter. 
2. Set the rotary switch to Hz. 
3. Set the Function Generator for the square wave voltage and frequency for steps 1-5 of Table 12. 
4. Compare the reading on the Meter display with the display reading shown in Table 12. 
5 If the display reading falls outside of the range shown in Table 12, the Meter does not 
 meet specification. 
Form Number TM61633-5  Rev 5 Dec 2006 