Thermometer User Manual
Page 16 
Testing the Capacitance Function (cont’d) 
 d) For each input, compare the readings on the Meter display to display readings in 
 Table 7. The meter selects the proper range automatically. Each measurement takes 
 about one second per range. 5mF will taken about 15 seconds. 
3. If the display reading falls outside of the range shown in Table 7, the Meter does not 
 meet specification. 
Table 7. Capacitance Test 
Input Readings 
Step Capacitance  61-633  61-635 
1  3.600nF  3.512 to 3.688  3.548 to 3.652 
2  36.00nF  35.12 to 36.88  35.48 to 36.52 
3  360.0nF  351.2 to 368.8  354.8 to 365.2 
4 3600μF  3.512 to 3.688  3.548 to 3.652 
5 36.00μF  34.76 to 37.24  35.12 to 36.88 
6 360.0μF  347.6 to 372.4  351.2 to 368.8 
7  1.000mF  .9410 to 1.059  .9510 to 1.049 
Checking the Diode Test Function 
To check the diode test function, do the following: 
1. Connect the Calibrator to the VΩ and COM inputs on the Meter. 
2. Turn the rotary switch to 
. 
3. Apply 3.000V. 
 The meter display should read approx. 3.000V dc. 
4. Press “Blue” key to test 
 function. 
5. Apply a 50Ω resistor to meter, the built-in beeper alarms. 
Testing the Milliamp (mA) Function 
To verify the accuracy of DC and AC current measurement functions, do 
The following: 
1. Connect the Calibrator to the mA and COM inputs on the Meter. 
2. Turn the rotary switch to mA. 
3. Apply the inputs for steps 1-2 in Table 8. 
4. For each input, compare the readings on the Meter display to the display readings in 
 Table 8. 
Table 8. DC mA Test 
Input Readings 
Step Current  61-633  61-635 
1  36.000rnA  35.780 to 36.220  35.888 to 36.112 
2  360.00mA  357.80 to 362.20  358.88 to 361.12 
5. Press the blue button on the Meter to toggle to AC measurement function. 
6. Apply the inputs for steps 1-4 in Table 9. 
7. For each input, compare the readings on the Meter display to the display readings in 
 Table 9. 
8. If the display reading falls outside of the range shown in the Tables, the meter does not 
 meet specification. 
Form Number TM61633-5  Rev 5 Dec 2006 