Thermometer User Manual
Page 15 
Testing the Resistance Function 
To verify the accuracy of the resistance function, do the following: 
1. Connect the Calibrator to VΩ and COM on the Meter. 
2. Turn the rotary switch to Ω 
3. Apply the inputs for step 1-7 in Table 5. 
 *Set reference if using a Multifunction Calibrator with diff or zero mode for steps 1, 2, and 3. 
 Compare the Meter display readings to the display readings in Table 5 step 1-7. 
4. If the display reading falls outside of the range shown in Table 5, the meter does not 
 meet specification. 
Table 5. Ω Test 
Input 
Readings 
Step 
Resistance Ω 
61-633 61-635 
1  0.00  00.00 to 00.02  00.00 to 00.02 
2  360.00  358.00 to 362.00  358.72 to 361.28 
3  3.6000K  3.5800 to 3.6200  3.5872 to 3.6128 
4  36.000K  35.800 to 36.200  35.872 to 36.128 
5  360.00K  358.00 to 362.00  358.72 to 361.28 
6  3.6000M  3.5780 to 3.6220  3.5852 to 3.6148 
7  36.000M  34.150 to 37.850  34.150 to 37.850 
5. Press the blue button on the meter to toggle to LVΩ function. 
6. Apply the inputs for step 1-6 in Table 6. 
 *Set reference if using a Multifunction Calibrator with diff or zero mode for steps 1 & 2. 
 Compare the Meter display readings to the display readings in Table 6. 
7. If the display reading falls outside of the reading shown in Table 6, the meter does not 
 meet specification. 
Table 6. LV Ω 
Input Readings 
Step 
Resistance Ω 
61-633 61-635 
1  0.00  00.00 to 00.02  00.00 to 00.02 
2  3.6000K  3.5620 to 3.6380  3.5764 to 3.6236 
3  36.000K  35.620 to 36.380  35.764 to 36.230 
4  360.00K  356.20 to 363.80  357.64 to 362.36 
5  3.6000M  3.5600 to 3.6400  3.5744 to 3.6256 
6  36.000M  33.430 to 38.570  33.430 to 38.570 
Testing the Capacitance Function 
The Meter measures capacitance by charging the capacitor with a known direct current, and measuring the 
resultant voltage, and calculating the capacitance. If the same capacitance is measured on an impedance bridge, a 
different reading may result. This variance is likely to be greater at higher frequencies. 
To verify the accuracy of the capacitance measuring function, do the following: 
1. Apply the Capacitor to the VΩ and COM inputs on the Meter. 
2. For steps 1 through 8 in Table 7. 
 a) Turn the rotary switch 
 b) Set the VΩ and COM inputs un-connected 
 c) Select Δ or (REL) for steps 1 and 2 in Table 7. After steps 1 and 2 are complete, escape 
 from Δ mode and press (RANGE) key until (AUTO) is displayed. 