(8) Capacitance
Range Resolution Accuracy
1.000μF 0.001μF
± (1.2% +2d)
10.00μF 0.01μF
100.0μF 0.1μF
1.000mF 0.001mF
10.00mF 0.01mF
Input Protection : 1000VDC or 1000VAC rms
Max. Measuring Time : 0.7 sec for 1nF~1mF
3 sec for 1mF~10mF
(9) Frequency Counter
Range Resolution Accuracy
100.00 Hz 0.1 Hz
± (0.1% +2d)
1000.0 Hz 0.1 Hz
10.000 KHz 0.001 KHz
100.00 KHz 0.01 KHz
Input Protection : 1000VDC or 1000VAC rms
Min. Frequency : 1Hz
Sensitivity : > 5.0Vp-p (for ACV 1HZ ~ 10kHz).
> 10Vp-p (for ACV 10kHZ ~ 100kHz)
> 2mAp-p (for ACmA)
> 0.2Ap-p (for ACA)
(10) Temperature (For 61-486 only)
Range Resolution Accuracy*
-40.0°C ~ 400°C 0.1°C ± (1.0% + 10d)
-40.0°F ~ 752°F 0.1°F ± (1.0% + 18d)
Input Protection : 1000VDC or 1000VAC rms
* Does not include accuracy of the thermocouple probe.
* Accuracy specification assumes surrounding temperature
stable to ± 1°C. For surrounding temperature changes of
± 5°C, rated accuracy applies after 2 hours.
Note : The AC+DC True RMS value might be over the
selected range since the AC signal may add on the DC level.