Use And Care Manual

17
18
(2) Auto-V
Range Resolution
For 61-484
Accuracy
For 61-486
Accuracy
600.0V
AC
0.1V
± (1.0% + 3d) ± (0.80% + 3d)
1000V
AC
1V
600.0V
AC
0.1V
1000V
AC
1V
Input Protection : 1000VDC or 1000VAC rms
AC Frequency Response : 50Hz ~ 1KHz
Input Impedance : Approx. 3kΩ.
AC Conversion Type :
Conversion type and additional specification are same as
voltage function.
(3) Current
Range Resolution
For 61-484
Accuracy
For 61-486
Accuracy
60.00mA
DC
0.01mA ± (1.0% + 3d) ± (0.08% + 3d)
600.0mA
DC
0.1mA ± (1.0% + 3d) ± (0.08% + 3d)
6.000a
DC
0.001A ± (1.0% + 3d) ± (0.08% + 3d)
10.00A
DC
0.01A ± (1.0% + 3d) ± (0.08% + 3d)
60.0mA
AC
0.01mA ± (1.5% + 3d) ± (1.2% + 3d)
600.0mA
AC
0.1mA ± (1.5% + 3d) ± (1.2% + 3d)
6.000A
AC
0.001A ± (1.5% + 3d) ± (1.2% + 3d)
10.00A
AC
0.01A ± (1.5% + 3d) ± (1.2% + 3d)
Input Protection :
Equipped with High Energy Fuse.
440mA, 1000V AC/DC 10KA Fuse for mA input.
11A, 1000V AC/DC 20KA Fuse for A input.
Max. Measuring Time : 3 minutes for A input, 10 minutes
for mA input.
Rest time 20 minutes min.
AC Conversion Type :
Conversion type and additional specification are same as
volt-age function.
AC Frequency Response : 50Hz ~ 1KHz