Installation Instructions
Page 21 
Testing the Temperature Function (for 61-340 / 61-342) 
To verify the accuracy of the meter’s temperature function, do the following: 
1.  Connect the calibrator to the 
 and COM inputs on the meter. 
2.  Set the rotary switch to the 
ºF / ºC position. Press the SELECT button to select the proper range as 
listed in Table 8. 
3.  Set the calibrators temperature output to the source for steps 1 - 7 in Table 8. 
4.  Compare the reading on the meter display with the display reading shown in Table 8. 
5.  If the display reading falls outside of the range shown in Table 8, the meter does not meet 
specification. 
Table 8 Temperature Test: 
Step  Range Source  Reading 
1  ºC -20 ºC  -23 to -17 
2  ºC 0 ºC  -3 to 3 
3  ºC 300 ºC  288 to 312 
4  ºC 750 ºC  725 to 775 
5  ºF -4 ºF  -9 to 1 
6  ºF 32 ºF  26 to 38 
7  ºF 1382 ºF  1336 to 1428 
Form Number TM61340-2  Rev 3 January 2008   