Installation Instructions
Page 20 
Testing the Frequency Function (for 61-340 / 61-342) 
      To verify the accuracy of the meter’s frequency function, do the following: 
1.  Connect the calibrator to the 
 and COM inputs on the meter. 
Note: The accuracy of the calibrator’s frequency function must be appropriate for the specified 
accuracy of the meter. 
2.  Set the rotary switch to the 
 position. 
3.  Set the calibrator or function generator for the square wave voltage and frequency for steps 1 - 6 of 
Table 7a. 
4.  Compare the reading on the meter display with the display reading shown in Table 7a. 
5.  If the display reading falls outside of the range shown in Table 7a, the meter does not meet 
specification. 
Table 7a Frequency Test: 
Step Range  Source  Level  Reading 
1  100Hz  9.0Hz  3V rms  8.988 to 9.012 
2 100Hz  90Hz 
3V rms 
89.88 to 90.12 
3 1kHz  900Hz 
3V rms 
898.8 to 901.2 
4 10kHz  9kHz 
3V rms 
8.988 to 9.012 
5 100kHz  90kHz 
3V rms 
89.88 to 90.12 
6 1MHz  900kHZ 
3V rms 
898.8 to 901.2 
7 10MHz 9.00MHZ 
5V rms 
8.97 to 9.03 
Testing the Duty Cycle Function (for 61-340 / 61-342) 
      To verify the accuracy of the meter’s frequency function, do the following: 
1.  Connect the calibrator to the 
 and COM inputs on the meter. 
Note: The accuracy of the calibrator’s frequency function must be appropriate for the specified 
accuracy of the meter. 
2.  Set the rotary switch to the 
  position. Press the Hz/DUTY button to select the DUTY CYCLE 
function 
3.  Set the calibrator or function generator for the square wave voltage and frequency in Table 7b. 
4.  Compare the reading on the meter display with the display reading shown in Table 7b. 
5.  If the display reading falls outside of the range shown in Table 7b, the meter does not meet 
specification. 
Table 7b Duty Cycle Test: 
Step Range  Source  Level  Reading 
1  1kHz  1kHz  3V rms  48.3% to 51.7% 