Testing the AC milliampere function: (for 61-340and 61-342) 
1.  Connect the calibrator to the 
 and COM inputs on the meter. 
2.  Turn the rotary switch to the 
 position. Press the SELECT button to select the AC function.   
Press the REL button to make sure the display is zero. 
3.  Apply the inputs for steps 1 - 4 in Table 4b. 
4.  For each input, compare the reading on the meter display to the reading for your meter in Table 4b. 
5.  If the display reading falls outside of the range shown in Table 4b, the meter does not meet 
specification. 
Table 4b ACmA Test:   
Step  Range  Source Frequency  Reading 
1  40 mA  30.00mA  60Hz  29.35 to 30.65 
2  40 mA  30.00mA  400Hz  29.35 to 30.65 
3  400 mA  300.0mA  60Hz  293.5 to 306.5 
4  400 mA  300.0mA  400Hz  293.5 to 306.5 
Testing the AC A function (for 61-340 / 61-342) 
1.  Connect the calibrator to the 10A and COM inputs on the meter. 
2.  Turn the rotary switch to the 
  position. Press the SELECT button to select the AC function. 
Press the REL button to make the display is zero. 
3.  Apply the inputs for steps 1 - 4 in Table 4c. 
4.  For each input, compare the reading on the meter display to the reading for your meter in Table 4c. 
5.  If the display reading falls outside of the range shown in Table 4c, the meter does not meet 
specification. 
Table 4c AC A Test:   
Step  Range  Source Frequency  Reading 
1  4A  3.000A  60Hz  2.920 to 3.080 
2  4A  3.000A  400Hz  2.920 to 3.080 
3  10A  9.00A  60Hz  8.73 to 9.27 
4  10A  9.00A  400Hz  8.73 to 9.27 