Installation Instructions
Page 16 
Table 3b DC mA Test:   
Step  Range Source  Reading 
1  40mA  30.00mA  29.80 to 30.20 
2  400mA  300.0mA  298.0 to 302.0 
Testing the DC ampere Function (for 61-340 and 61-342) 
1.  Connect the calibrator to the 10A and COM inputs on the meter. 
2.  Turn the rotary switch to the 
 position. Press the SELECT button to select the DC function 
3.  Apply the inputs for steps 1 - 2 in Table 3c. 
4.  For each input, compare the reading on the meter display to the reading for your meter in Table 3c. 
5.  If the display reading falls outside of the range shown in Table 3c, the meter does not meet 
specification. 
Table 3c DCA Test:   
Step  Range Source  Reading 
1  4A  3.000A  2.950 to 3.050 
2  10A  9.00A  8.82 to 9.18 
Testing the AC microampere Function (for 61-340 and 61-342) 
To verify the accuracy of AC and DC current measurement functions, do the following: 
1.  Connect the calibrator to the 
 and COM inputs on the meter. 
2.  Turn the rotary switch to the 
 position. Press the SELECT button to select the AC function.   
Press the REL button to make sure the display is zero. 
3.  Apply the inputs for steps 1 - 6 in Table 4a. 
4.  For each input, compare the reading on the meter display to the reading for your meter in Table 4a. 
5.  If the display reading falls outside of the range shown in Table 4a, the meter does not meet 
specification. 
Table 4a AC microampere Test:   
Step  Range Source  Frequency Reading 
1  400µA  40.0µA  60Hz  38.9 to 41.1 
2  400µA  40.0µA  400Hz  38.9 to 41.1 
3  400µA  300.0µA  60Hz  295.0 to 305.0 
4  400µA  300.0µA  400Hz  295.0 to 305.0 
5  4000µA  3000µA  60Hz  2950 to 3050 
6  4000µA  3000µA  400Hz  2950 to 3050 
Form Number TM61340-2  Rev 3 January 2008   