Digital Photo Keychain User Manual
Form Number TM61310  Rev 2 Aug 2007 
Page 11 
e Test:   Table 2: AC Voltag
Step 
Range 
Input  Frequency
  Reading 
1 
200V 
2.0V  60Hz  1.0 to 3.0 
2 
200V 
2.0V    400Hz  1.0 to 3.0 
3 
200V 
190.0V  60Hz  186.7 to 193.3 
4 
200V 
190.0V  400Hz  186.7 to 193.3 
5 
600V 
600V  60Hz  583.0 to 617.0 
6 
600V 
600V  400Hz  583.0 to 617.0 
Testing the DC milliamp
1
s 1-3 in Table 3. 
meter 
r does not meet 
s and DC amps Function   
.  Turn the rotary switch to the DCA function and range shown in Table 3. 
2.  Apply the inputs for step
3.  For each input, compare the reading on the meter display to the reading for your 
in Table 3. 
4.  If the display reading falls outside of the range shown in Table 3, the mete
specification. 
  Table 3: DC Current Test:   
Step 
Range 
Source    Reading 
1 
200 mA 
2.00 mA  1.8 to 2.2 
2 
200 mA 
190.0 mA  186.9 to 193.1 
3 
10 A 
10.00 A  9.65 to 10.35 
Testing the Resis ce Fun
To verify the accur  of the  e function  following: 
1.  Connect the calibrator to  and COM on the meter. 
2.  Turn the rot  switch to OHM. function  e shown
3.  Apply the inputs for step Table 4.
4.  Compare th eter disp adings to t y readin
5.  If the display reading falls out e of the range shown in Table 4, the  s not meet 
specification. 
tan ction 
acy resistanc , do the
 V
ary and rang  in Table 4   
s 1-6 in   
e m lay re he displa gs in Table 4. 
sid meter doe