User Manual
7
5.
Self-tuning of PID
When the temperature control effect is not ideal, a system self-tuning can be carried out where the
equipment will automatically configurate its PID parameters. To do that, follow the next instructions:
-
Press
AT
for 6s
the controller will run the auto-tuning function. The
AT
indicator will flash.
-
To stop the process, press
AT
for 6s while it is running.
-
During the process, the
SET
button does not work, and the lower window always displays the
temperature set value.
6.
Internal parameters setting
6.1.
How to enter
To adjust the internal parameters, you must enter the required password. To do that, follow these
instructions:
1.
Press
SET
for 3s
the controller will display the password prompt
Lc
.
2.
Use
◄ ▼▲
to choose the required value (see the password values below).
3.
Press
SET
to enter
the controller will enter the internal parameters setting state.
To return to the regular state at any time, press SET for 3s.
6.2.
Parameter tables
INDICATOR
NAME
INSTRUCTIONS OF THE FUNCTION
(RANGE)
DEFAULT VALUE
Lc
Password
The password
to modify
the
parameters
in this
table is
Lc = 3
0
ALH
Over-temp
alarm
If
SV > (SP + ALH)
, the
ALM
light turns on. The buzzer
sounds and the heating output turns off.
(0
~
100.0
℃
)
20.0
ALL
Under-temp
alarm
If
SV < (SP – ALL)
, the
ALM
light flashes, the buzzer
sounds.
(0
~
100.0
℃
)
0
T
Control cycle
The temperature control cycle.
(1
~
60S)
5
P
Proportional
band
Adjustment of proportional function.
(1
~
400.0
℃
)
35.0
I
Integration
time
Adjustment of integration function.
(1
~
2000S)
200
d
Differential
time
Adjustment of differential function.
(0
~
1000S)
200
Pb
Zero point
adjustment
When
the
error
is
comparatively
larger,
this
value
might
need adjustment.
Pb =
actual value
–
measure value
(-12.0
~
12.0
℃
)
0
PK
Large value
adjustment
Correction
of errors
caused
by
sensor
high
temperature
values.
actual value
-
measured value
PK
=
1000
×
measured value
(-999
~
999)
0