User's Manual

www.i-sens.com 24
You may check your meter and test strips using the AutoChek t Control
Solution.
The AutoChek t Control Solution contains a known amount of glucose
and is used to check that the meter and the test strips are working
properly.
The test strip vials have AutoChek t Control Solution ranges printed on
their labels. Compare the result displayed on the meter to the AutoChek
t Control Solution range printed on the test strip vial.
Before using a new meter or a new vial of test strips, you may
conduct a control solution test following the procedure on page 18.
Note:
Use only the AutoChek t Control Solution.
Check the expiration dates printed on the vial. When you first open a
control solution vial, record the discard date (date opened plus three
months) in the space provided on the label.
Make sure your meter, test strips, and control solution are at room
temperature before testing. Control Solution tests must be done at room
temperature (20 ~ 25˚C, 68 ~ 77˚F).
Before using the control solution, shake the vial, discard the first few drops
and wipe the tip clean.
Close the control solution vial tightly and store at
a temperature of 8 ~ 30˚C (46 ~ 86˚F).
You may do a control solution test:
When using the meter for the first time.
Whenever you open a new vial of test strips.
If the meter or test strips do not function properly.
If your symptoms are inconsistent with the blood glucose test results, and when
you feel that the meter or test strips are not working properly
If you drop or damage the meter.
Checking the System