Message What It Means What To Do
The blood sample
has abnormally
high viscosity or
insufficient volume.
Repeat the test after
inserting a new test strip.
A non-CareSens
LINK test strip was
used.
Test strip was not
inserted properly.
Repeat the test with a
CareSens LINK test strip.
Insert a test strip with the
contacting bars facing
upwards and push in gently
until the meter beeps.
There is a problem
with the meter.
Do not use the meter.
Contact your authorized
i-SENS sales
representative.
If the error messages persist, contact your authorized i-SENS
sales representative.
Note
General Troubleshooting
Problem Troubleshooting
Check whether the test strip is
inserted with the contact bars facing
up. Check if the strip has been
inserted completely till the end.
Check whether the batteries are
inserted with the+’ side facing up.
Replace the batteries.
The display is blank
even after inserting a
test strip.
Check if the confirmation window is
filled adequately.
Repeat the test after inserting a new
test strip.
The test does not start
even after applying the
blood sample on the
strip.
Repeat the test after inserting a new
test strip.
Check the validity period of the test
strip.
Check if the strip has passed the
discard date (date of opening the test
strip vial plus 6 months).
Check the meter.
The test result
doesn’t match your
expectation.
If the problem is not resolved, please contact your authorized
i-SENS sales representative.