User's Manual
www.i-sens.com
42
www.i-sens.com
43
Caring for Your System
Use a soft cloth or tissue to wipe the meter exterior. If necessary, the
soft cloth or tissue might be dipped in a small amount of alcohol.
Do not use organic solvents such as benzene, or acetone, or
household and industrial cleaners that may cause irreparable damage
to the meter.
Store all the meter components in the portable case to prevent loss.
• Do not expose the meter to direct sunlight or heat for an extended
period of time.
• Prevent the entry of dirt, dust, blood, or water at the meter's test
port.
•Do not drop the meter or submit it to strong shocks.
•Do not try to fix or alter the meter in any way.
• Keep the meter away from strong electromagnetic fields such as
cell phones and microwave ovens.
• CareSens LINK meter should be used only with CareSens LINK
strips.
Caution:
Understanding Error and Other Messages
Message What It Means What To Do
A used test strip
was inserted.
Repeat the test with a
new test strip.
The blood or
control solution
sample was
applied before
the
icon
appeared.
Repeat the test with
a new test strip and
wait until the icon
appears before applying
the blood sample.
The temperature
during the test was
above or below the
operating range.
Move to an area where
the temperature is within
the operating range (10
~ 40˚C/ 50 ~ 104˚F) and
repeat the test after 30
minutes.