User's Manual

www.i-sens.com
22
Repeat the control test if the test result falls outside the range printed
on label of the test strip vial. Out of range results may occur due to the
following factors:
Comparing the Control Solution Test Results
When the control solution vial was not shaken well,
When the control solution is past its expiration date or is
contaminated,
When the meter, the strip or the control solution were exposed to
high or low temperatures,
When the first few drops of the control solution were not discarded
or the tip of the vial was not wiped clean,
When the test strip is past its expiration date,
When the meter is not functioning properly.
The CareSens Control Solution can be purchased separately.
Please contact your authorized i-SENS sales representative.
Note
www.i-sens.com
23
Using the Lancing Device
You will need a lancing device in order to collect a blood sample.
You may use the lancing device contained in the CareSens LINK
Blood Glucose Monitoring System or any other medically approved
lancing device.
The lancing device should be used by one individual and communal
use is strongly discouraged.
Use a soft cloth or tissue to wipe the lancing device. If necessary, a
small amount of alcohol on a soft cloth or tissue may be used.
To avoid infection when drawing a sample, use a lancet only one
time, and:
Do not use a lancet that has been used by others.
Always use a new sterile lancet.
Keep the lancing device clean.
Caution
Repeated puncturing at the same sample site may cause pain or
skin calluses. Choose a different site each time you test.
Note
Adjustable Tip
Lancet Holder
Sliding Barrel
Release Button
Ejection System