User's Manual
Table Of Contents
28 29
CONTROL SOLUTION TESTING
Step 4:
The test result will appear after the meter counts
down from ve (5) to one (1) in 5 seconds.
You can see the control solution mark “Check” on
screen. In this case, the result is stored in the meter
memory but not included in the averages.
Step 5
Compare the result with the control solution
range printed on the Mendor SMART Blood Glucose
Meter Test Strip bag. The result should fall within that
range. Used strips should be discarded safely in dis-
mmol/L
6.7
CONTROL SOLUTION TESTING
Comparing the Control Solution Results
If your control solution test results do not fall within the range printed on the
test strip vial, repeat the test. Out of range results may occur due to the following
factors:
• When the control solution bottle was not shaken well
• When the control solution is past its expiration date or is contaminated
• When the meter, the strip or the control solution were exposed to high or
low temperatures
• When the rst few drops of the control solution were not discarded, or the tip
of the vial was not wiped clean
• When the test strip is past its expiration date
• When the meter is not functioning properly
NOTE! If you cannot perform an acceptable control measurement (that falls
within the control range) do not use the blood glucose monitoring system and
contact the manufacturer of the blood glucose meter.