User manual

the input accurately and the display changes from one reading to
the next at the voltage half way between the two values. Hence,
with dropping voltage the display will change at
4.95 V from 5.0 to 4.9V.
Lower-limit calibration. This is what you may choose to set if you
want the voltage alarm to trigger as soon as that voltage is reached.
Using this approach, at 4.90 V the display will change
from 5.0 to 4.9V. In practice, flight pack and receiver voltage alarms
are intended to alert you to when the voltage has dropped to a critical level. Lower-limit calibration
will do that; whereas mid-point calibration will not trigger the alarm until the voltage has dropped by
another 0.05Volts.
Bottom line: Use lower limit calibration to achieve most accurate alarms.
Setting lower-limit calibration is easy. For example, with a known voltage of 5.00V, adjust the
calibration until it is just flickering between 5.0 and 4.9V. Mid-point calibration is set by putting the
pot halfway between where the display flickers at the lower and
upper value. The values used are just examples for illustration, and
the principles above apply to whatever actual voltage you are
calibrating against.
Product Note:
This is not a Horizon Hobbies DSM2™, DSMX™ product, and is not manufactured or
endorsed by Horizon Hobbies USA. DSM2™, and DSMX™ are registered trademarks of
Horizon Hobbies LLC.
Accepted by the MAAA in Australia, see the MAAA MOP58 for guidance
www.Hyperion-World.com