Manual

User Manual HYDACLab
23.10.2008
8
6.3 Reset using HMG 3000
The HYDACLab
can be reset in the sensor dialogue of the HMG 3000 (for connection of the
HYDACLab
to a HMG 3000, see Chap. 12.3 HMG 3000).
6.4 Reset using PC software CMWIN
The HYDACLab
can be reset in the sensor dialogue of CMWIN.
CMWIN is special HYDAC PC software and is supplied with the
HMG 510 and HMG 3000.
7. Electrical connection
Signal 1
HYDACLab
2
1
3
4
+U
B
Signal 2
RL
5
GND
Signal 1: PNP-Schaltausgang
Signal 2: Sequentieller Analogausgang (4 .. 20 mA)
HSI
*
/ Reset (SPS)
PIN 5: HSI-Kanal / Reset (SPS)
R
Lmax
= (U
B
-5V) / 20mA [k] and R 500
Note:
The load resistance R
L
is the sum of the internal input resistance of the evaluation unit and the resistance of
the connection line.
* HSI = HYDAC Sensor Interface (HYDAC's own communication interface)