Instruction Manual

ETS 3000 User Manual 05 Sep 2008
6
6.2 Analogue Output
Analogue output signal: 4 – 20 mA or 0 – 10 V
(can be set in the Basic Settings menu)
6.3 Setting the Switching Points and Hysteresis or Switching Values for Window
Mode
Press the mode key.
S.P.1 or Lo.1 appears in the display.
Press the mode key repeatedly to scroll through to the required parameter.
(S.P.1, hYS.1, S.P.2 or hYS.2 appears when switching point mode is set; Hi.1, Lo.1, Hi.2
or Lo.2 appears when window mode is set)
After 2 seconds the actual setting will flash.
Change the setting using the
and keys.
If necessary, use the mode key to select further parameters, and the
and keys to
change the setting.
The display switches back automatically if no keys have been pressed for 3 s; the settings
are saved.
2 s
3 s
Press the mode key
Display
Adjust the switching point:
until the required
parameter is reached
=decreases value
=increases value
SP1
SP2
HYDAC
mode
°C
*
SP1
SP2
HYDAC
mode
*
°C
°C
°C
*
Actual
switching point 1
SP1
SP2
HYDAC
mode
°C
Notes:
If "LOC" appears in the display while attempting to change a setting, this means that
programming is disabled (locked). Corrective action:
Set the program enables to "free".
(see section 8 Programming Enables))
If the
or key is held down while making a change, the value will automatically
advance.
When a setting has been changed,
PRoG appears briefly in the display when it switches
over to another display mode. The new setting has then been saved.