User guide

ContaminationSensor CS1000 Analog Output (ANAOUT)
HYDAC Filtertechnik GmbH
en
Page 74 / 100
BeWa CS1000 3247149n en 2009-02-04.doc 2009-02-04
HDA.NAS – Analog Signal NAS to HDA 5500
The HDA.NAS signal consists of 4 measured values (Start / NAS 2 / NAS 5 / NAS
15 / NAS 25 / Status) which are output sequentially. Synchronization with the
downstream control unit is a prerequisite.
The signal output is as follows:
Time Measurement
category
Signal duration
in s
Current/Voltage
Start signal 0 -- 2 20 mA / 10 V
Pause 2 4 mA / 2 V
Signal 1 2-5 µm 2 Current/Voltage for signal
Pause 2 4 mA / 2 V
Signal 2 5-15 µm 2 Current/Voltage for signal
Pause 2 4 mA / 2 V
Signal 3 15-25 µm 2 Current/Voltage for signal
Pause 2 4 mA / 2 V
Signal 4 > 25 µm 2 Current/Voltage for signal
Pause 2 4 mA / 2 V
Signal 5 Status 2 Current/voltage for signal
Pause 30 4 mA / 2 V