User guide

ContaminationSensor CS1000 Analog Output (ANAOUT)
HYDAC Filtertechnik GmbH
en
Page 65 / 100
BeWa CS1000 3247149n en 2009-02-04.doc 2009-02-04
ISO 2 / ISO 5 / ISO 15 (ISO 2 / ISO 5 / ISO 15)
The ISO x setting enables the value of a class to be continuously output via the
analog output.
ISO code (ISO), 3-digit
The ISO code signal consists of 3 measured values (>2 µm / >5 µm / >15 µm) which
are copied time-coded as shown below.
4,0
4,1
I (mA)
1
3
2
4
6
2
5
1
U (V)
t (ms)
4,8
19,2
High
High
Low
Low
300
3000
3000
3000
9,75
19,5
2,15
2,05
2,0
9,85
19,7
0,0
19,8
9,9
Time
Signal Size Signal duration
per pulse
in ms
Current (I) / Voltage (U)
1 Identifier > 2 µm 300 High / Low
2 Measured value > 2 µm 3000 Current/Voltage for measured value
3 Identifier > 5 µm 300 High / Low / High / Low
4 Measured value > 5 µm 3000 Current/Voltage for measured value
5 Identifier > 15 µm 300 High / Low / High / Low / High / Low
6 Measured value > 15 µm 3000 Current/Voltage for measured value