User guide

ContaminationSensor CS1000 Analog Output (ANAOUT)
HYDAC Filtertechnik GmbH
en
Page 63 / 100
BeWa CS1000 3247149n en 2009-02-04.doc 2009-02-04
ISO code signal acc. to 4406:1987 (CS 13xx only)
The following ISO values can be output via the analog output:
ISO 2 / ISO 5 / ISO 15
Only one value is output.
ISO-Code, 3-digit (>2 µm / >5 µm / >15 µm)
All values are sequentially time-coded before output.
ISO+T
All values are sequentially time-coded before output.
HDA.ISO
All values are sequentially time-coded before output.
Primarily this signal was planned for the HDA 5500 (HYDAC digital display Unit),
however, this can use in other applications.
The current 4.8 - -19.2 mA or voltage 2.4 - 9.6 V of the output signal is dependent
on the ISO Contamination Class 0.0 - 24.28 (Resolution 1 Class) or an error as
shown in the table below.
Current I ISO contamination class / error Voltage U
I <4.00 mA Cable break U <2.00 V
4.0 mA < I < 4.1 mA Device error, device not ready 2.0 V < U < 2.05 V
4.1 mA < I < 4.3 mA Not defined 2.05 V < U < 2.15 V
4.3 mA < I < 4.5 mA Flow error (flow insufficient) 2.15 V < U < 2.25 V
4.5 mA < I < 4.8 mA Not defined 2.25 V < U < 2.4 V
I = 4.80 mA ISO 0 U = 2.40 V
I = 5.37 mA ISO 1 U = 2.69 V
I = 5.94 mA ISO 2 U = 2.97 V
I = 6.51 mA ISO 3 U = 3.26 V
I = 7.08 mA ISO 4 U = 3.54 V
I = 7.65 mA ISO 5 U = 3.83 V
I = 8.22 mA ISO 6 U = 4.11 V
I = 8.79 mA ISO 7 U = 4.40 V
I = 9.36 mA ISO 8 U = 4.68 V
I = 9.93 mA ISO 9 U = 4.97 V
I = 10.50 mA ISO 10 U = 5.25 V
I = 11.07 mA ISO 11 U = 5.54 V
I = 11.64 mA ISO 12 U = 5.82 V
I = 12.21 mA ISO 13 U = 6.11 V
I = 12.77 mA ISO 14 U = 6.39 V
I = 13.34 mA ISO 15 U = 6.67 V
I = 13.91 mA ISO 16 U = 6.96 V
I = 14.48 mA ISO 17 U = 7.24 V
I = 15.05 mA ISO 18 U = 7.53 V
I = 15.62 mA ISO 19 U = 7.81 V
I = 16.19 mA ISO 20 U = 8.10 V
I = 16.76 mA ISO 21 U = 8.38 V