User guide

ContaminationSensor CS1000 Analog Output (ANAOUT)
HYDAC Filtertechnik GmbH
en
Page 58 / 100
BeWa CS1000 3247149n en 2009-02-04.doc 2009-02-04
ISO 4 / ISO 6 / ISO 14 (ISO 4 / ISO 6 / ISO 14)
The ISO x setting enables the value of a class to be continuously output via the
analog output.
ISO code (ISO), 3-digit
The ISO code signal consists of 3 measured values (>4µm
(c)
/ >6µm
(c)
/ >14µm
(c)
)
and is time-coded with the following time slices.
4,0
4,1
I (mA)
1
3
2
4
6
2
5
1
U (V)
t (ms)
4,8
19,2
High
High
Low
Low
300
3000
3000
3000
9,75
19,5
2,15
2,05
2,0
9,85
19,7
0,0
19,8
9,9
Time
Size Signal duration
per pulse
in ms
Current (I) / Voltage (U)
1 Identifier > 4 µm(c) 300 High / Low
2 Measured
value
> 4 µm(c) 3000
Current/Voltage for measured value
3 Identifier > 6 µm(c) 300 High / Low / High / Low
4 Measured
value
> 6 µm(c) 3000
Current/Voltage for measured value
5 Identifier > 14 µm(c) 300 High / Low / High / Low / High / Low
6 Measured
value
> 14 µm(c) 3000
Current/Voltage for measured value