User guide
ContaminationSensor CS1000 Analog Output (ANAOUT)
HYDAC Filtertechnik GmbH
en
Page 52 / 100
BeWa CS1000 3247149n en 2009-02-04.doc 2009-02-04
SAE A / SAE B / SAE C / SAE D (SAE A/SAE B/SAE C/SAE D)
The SAE x setting enables the value of a class to be continuously output via the
analog output.
SAE + T (SAE+T)
The SAE+T signal consists of 5 measured values which are transmitted time-coded
with the following time slices:
4,0 2,0
I (mA)
1
2
4
6
8
10
3
5
7
9
U (V)
time (ms)
4,8
19,2
High
High
Low
Low
300
3000
3000
3000
3000
2,25
2,4
9,6
4,5
9,7519,5
9,9
10,0
9,85
19,7
0,0
19,8
Time
Signal Size Signal
duration
per pulse
in ms
Current (I) / Voltage (U)
1 Identifier SAE A 300 High / Low
2 Measured
value
SAE A 3000
Current/Voltage for measured value
3 Identifier SAE B 300 High / Low / High / Low
4 Measured
value
SAE B 3000
Current/Voltage for measured value
5 Identifier SAE C 300 High / Low / High / Low / High / Low
6 Measured
value
SAE C 3000
Current/Voltage for measured value
7 Identifier SAE D 300 High / Low / High / Low / High / Low / High / Low
8 Measured
value
SAE D 3000
Current/Voltage for measured value
9 Identifier Temperature 300 High / Low / High / Low / High / Low / High / Low /
High / Low
10 Measured
value
Temperature 3000
Current/Voltage for measured value