Owner manual

Using a ½" hex socket on a
3
/
8
" drive ratchet, connect to the
internal male hex shaft and crank counter clockwise to raise the
platen and apply load to the reference gage and the machine’s
load cell.
Starting from the chart origin, observe the recorder's pen position
and check it against the reference gage reading for each 1000 lbf
of load applied.
If an adjustment other than a zero shift is needed, remove the
recorder from the chassis leaving the cables connected and from
the picture below find potentiometer R231 (for the Y axis). With
a small, plastic screw driver turn the pot to increase or decrease
the gain as required.
Calibration of the Flow or X axis is accomplished by measuring the platen
travel with a rule while keeping time with a stop watch (usually
1
/
2
" in 15
seconds is the measured distance and time check) while observing the chart
reading. If the travel is not within .50" +/- .04" on the chart adjust the gain
using pot R131.
Maintenance
The machine and transducers should be kept clean and the machine should not be
over lubricated. Light oiling with synthetic instrument oil at most is required on
exposed spindles and threads; the jack and gear box are serviced with a premium
wheel bearing grease such as Pennzoil 707L Lubriplate Grease applied thru the
grease fitting no more than annually, or when an indication of power train friction
appears.
R231
R131
Top End
6