Owner manual

Physical Specification
Net weight: 226 lbs. (110 kg)
Shipping weight: 266 lbs. (113.6 kg)
Overall height: 39"
Base dimension: 35"x18"x10"
Power Supply
110 VAC 60 Hz 1 phase (H-1322B), 975 watts
220 VAC 60 Hz 1 phase (H-1322B.2F), 975 watts
220 V AC 50 Hz 1 phase (H-1322B.5F), 975 watts
Check that your machine has the correct voltage for your local supply.
Power Cord
Your Marshall & TSR Test Machine is fitted with a fused IEC receptacle and
cord with molded vinyl-grounding plug. The cord is made of three conductors:
(American color-coding)
Green Ground (Earth)
White Cold (Neutral)
Black Hot (Live)
or
(European color coding)
Green /Yellow Ground (Earth)
Blue Cold (Neutral)
Brown Hot (Live)
Fuses
The Marshall Test Machine uses a slo-blow fuse that is fitted into the rear of
the machine. The replacement part a 5 x 20 mm 10A, 250V ceramic (or glass).
Initial Installation
Remove the packing list and verify that all parts have been received. Normally you
will receive one crate comprised of compression and recorder frame, with calibrated
load cell (HM-2300.100) 10,000lbf capacity (44kN) fastened to it, 100 charts
(H-1322CP), 6 pens, a DB9 serial cable, and this user manual.
Set the machine base up at the place it is to be used in the laboratory and then
check that it is level by placing a spirit level on the platen. If it is not level, adjust the
feet of the machine.
Operation
Control Panel
The control panel is situated on the front right hand side of the machine base.
Recorder range and pen controls are on the recorder's chart plate.
Preset Speed of Platen:
2.00 in/min. (50.80mm/min.)
3