Operating Instructions

180°
14
7. Calibration and adjustment 8. Maintenance, storage and
transportation
Calibration of gradient module
Calibration stage
1
based on position 0°.
Place the DPL with the measurement surface
horizontally.
Briefly press buttons and .
Display -1- appears on the screen.
Press the button briefly, display -1- flashes
for approx. 5 seconds.
Display -2- appears on the screen.
Turn DPL again by 180°.
Press the button briefly, display -2- flashes
for approx. 5 seconds.
Measurement display appears on the screen,
calibration stage 1 is complete.
8.1 Cleaning
Wipe o the dirt with a so damp cloth.
Do not use aggressive cleaning agents or
solvents.
Do not immerse the device into water!
Clean and dry wet equipment, accessories
and transport containers prior to packaging
them. Only pack equipment again when it is
completely dry.
Keep plug connections clean and protected
from moisture.
8.2 Storage
8.2.1 GENERAL
e equipment may only be stored within the
specified temperature limits
(see Chapter 3 Technical data).
Aer prolonged storage check the accuracy
of the measuring device before using it.
8.2.2 BATTERIES/RECHARGEABLE BATTERIES
To store the rechargeable battery remove it
from the device or the charging station.
e storage should preferably be in a dry
environment at room temperature
(see Chapter 3 Technical data).
Protect from moisture and humidity. Dry wet or
damp batteries before storage, or before usage.