Owner manual

13
Figure 2. 4.1
Volume This is the total volume that will be primed through the system. Enter a
value between 120 and 9999uL.
Speed This adjusts the percentage of the pumps maximum speed. More viscous
liquids will require a slower pump speed to avoid cavitation. Do not exceed more
than 30% speed on any value over 1000ul.
Tip Touch Activate or deactivate tip touching after each dispense.
Tip Touch Adjust where in the Priming trough you would like to touch the tips
after a prime. This is helpful when running viscous liquids.
2.5 - Empty Dispenser
This feature reverses the pump, removing most of the liquid from the manifold
and tubing and returns it to the source container (Figure. 2.5 1).