User Manual
HT9019
EN - 14
6.1.1. Safety
Comply with: IEC/EN61010-1
Insulation: double insulation
Pollution degree: 2
Max height of use: 2000m (6562 ft)
Installation category: CAT IV 600V, CAT III 1000V to ground,
6.1.2. General data
Mechanical characteristics
Dimensions (L x W x H): 252 x 88 x 44mm ; 10 x 3 x 2 in
Weight (including battery): 402g (14 ounces)
Max conductor size: 45mm (2 in)
Supply
Battery type: 1x9V battery NEDA 1604 IEC 6F22 JIS 006P.
Low battery indication: “
” is displayed when the battery level is low
Battery life: about 200 hours
Auto Power OFF: after 15 minutes of idleness (not disabled)
Display
Characteristics: 4 LCD (max 6000 counts), decimal point, unit symbol
indication, bargraph and backlight
Sample rate: 2 times/sec
Conversion mode: TRMS
6.2. ENVIRONMENTAL CONDITIONS
6.2.1. Climatic conditions
Reference temperature: 18°C 28°C (65°F 83°F)
Operating temperature: 5 ÷ 40 °C (41°F 104°F)
Operating humidity: < 75%RH
Storage temperature: -20 ÷ 60 °C (-4°F 140°F)
Storage humidity: < 80% RH
This product conforms to the prescriptions of the European directive on low voltage
2006/95/EEC and to EMC directive 2004/108/EEC
6.3. STANDARD ACCESSORIES
Test leads – Code 4413-2
Carrying bag
User manual
Battery