User Manual

EN - 12
Non
-
contact forehaed infrared thermometer
11.Settings
11-1. Unit selection – Function F1
1. Press
MODE
key for 2 seconds, the screen displays “F1”
2. Press
DOWN key for degrees Celsius or UP key for degrees
Fahrenheit
11-2. Alarm setup – Function F2
1. Press
MODE
key for 2 seconds, the screen displays “F1”
2. Press t wice
MODE
touch to get “F2” . Press
UP
key to increase the
threshold by 0
.1°C (0.1°F)
or
DOWN
key
to reduce it by 0.1°C
(0.1
°F).
NOTE
: The alarm threshold default value is 38°C (100.4°F)
11-3. Temperature adjustment – Function F3
To adjust the total variation of your the Non-contact Body lnfrared
Thermometer:
1.
Press
MODE
key for 2 seconds, the screen displays “F1”
2.
Press
"MODE"
button two times until shows the “F3”
3.
Press
UP"
to increase the difference by 0.1°C (0.1 °F),
"DOWN"
to
reduce it by 0.1
°C (0.1 °F)
4.
ln case of seasonal or environmental changes a verification and
adjustment should be carried out.
NOTA:
This function is only effective Body
11-4. Buzzer ON/OFF - F4 Menu
1.
Press
MODE
key for 2 seconds, the screen displays “F1”
2.
Premere il tasto MODE fino a visualizzare il messaggio “F4”
3.
Press
UP
key to activate the buzzer (a sound icon ”
is
displayed on the LCD screen) or press
DOWN
key
to disable it
(the icon ” ” disappear)
11-5. Exit Setting mode
Press
MODE key
until the screen turns off