User Manual
EN - 10
Non
-
contact forehaed infrared thermometer
9.
Initial calibration
For stable and reliable results, The thermometer provides a user-correction
function, the operation of the method is as follows:
1
st
step
take the temperature of a person using a conventional
thermometer; you will get 37.5
°C (99.5°F) for instance
2
nd
step
take the temperature of the same person using the Non-
contact Body lnfrared Thermometer keeping the 5 to 15 cm (2 in to
5.9 in) distance between the thermometer and the forehead (Take
care to remove any obstacle which could alter the measurement (hair
,
perspiration ...). lf you get 37.5°C (99.5 °F), the Non-contact Body
lnfrared Thermometer is properly set and ready for use
lf you get a lower temperature, 36.4°C (97 .4°F) for example, your
difference is 1.1°C (2 .2°F). You should adjust the temperature on the
Non-contact Body lnfrared Thermometer and add the difference, i.e.
1.1°C (2 .2°F)
:
Press
MODE
key for 2 seconds, the screen displays “F1”
Press
MODE
key again until you get “F3”
Press
UP
key in order to add the difference (in our example,
1.1
°c - 2.2°F)
To check, take the temperature again using the Non-contact
Body mode