7.1. ENVIRONMENTAL PARAMETER AND LEAKAGE CURRENT: REAL TIME
MEASUREMENT
This working mode allows to execute real time measurement and recording of
Environmental Parameter and Leakage current
1. Press this key to access to "AUX" mode.
2. Pressing these function keys you will change the measuring unit of the 1st, 2nd,
3rd instrument's input respectively. The following possibilities will be displayed
circularly:
- - - (Instrument's input disabled)
mA (Leakage current)
°C (Celsius Temperature)
°F (Fahrenheit Temperature)
HR% (Relative humidity)
m/s (Air Speed)
mV (Voltage)
LUX (20) (Illuminance: Full Scale 20Lux)
LUX (2k) (Illuminance: Full Scale 2kLux)
LUX (20k) (Illuminance: Full Scale 20kLux)
3. Connect the External probes or clamps to the corresponding instrument inputs.
4. Check if the selector on the probes or clamps is adjusted according to the range
set in the instrument. Both range must be set to the same value.
B1
B2 B3 B4
I3
I2
I1
Example of External probes connections
WARNING
The OFF position of few probes send to probe's output the Battery voltage
(approx 9V which is over the expect full scale). This could influence the
measurement of the other instrument's inputs. So NEVER left connect to
instrument probes with the Selector placed in OFF position.