Manual
GSC53N - GSC57 - ZG47
EN - 63
6.6.1. Measurement Procedure and Result of "RMEAS" mode
1. Select the "RMEAS" mode by means of the F1 key.
2. Connect the test leads to B1, B2, B3, B4 input terminals according to the
following picture:
B2 B3
B4B1
I1
I2
I3
230V~
50Hz
1.3A
Connection of the test leads during LOW 10A test.
3. Set the resistance limit value by mean the F3 and F4 keys.
4. Connect the 230V~ 50Hz Power supply socket of the instrument to a power socket
using the C5700 cable.
5. Press the START key. The instrument will execute the measurement..
WARNING
If the instrument displays "Measuring" it means that it is effecting the
measurement. During this phase the instrument test leads are not to be
disconnected.
At the end of the test the
instrument emits an
acoustic signal and
display the screen
alongside.
LOW 10A 05.06.01
Resistance value
0.107
13.4A
RMEAS RLIM:0.150
Test Current Value
FUNC
Working mode
The displayed result can be stored pressing the SAVE key twice (according to
paragraph 9.1).
START
STOP