Manual

GSC53N - GSC57 - ZG47
EN - 39
During the ramp test if
the RCD tripping current
is higher than I
n
(Type
AC) or 1.4 I
n
(Type A
with I
n
>10mA) or 2 I
n
(Type A with I
n
10mA),
the instrument emits a
long sound signal at the
end of the test and
displays the values
alongside.
RCD 05.06.01
Maximum current generated by
the instrument during the test for
general RCDs (the value
indicated is referred to an AC
type 30mA RCD, in this case the
maximum current supplied is
equal to 1.4xI
N
> 42mA
>300ms
FRQ=50.0Hz Ut= 1V
VP-N=231V VP-PE=230V
CURRENT NOT OK
30mA 50V
ATTENTION:
the RCD tripping current is higher
than the Nominal Value
(I
N
=30mA was set in the
example).
FUNC
IdN
RCD
UL
The test can be stored pressing the SAVE key twice (according to paragraph 9.1).