Manual

GSC53N - GSC57 - ZG47
EN - 24
6.2.3. Results of "TMR" mode
At the end of the test if
the Insulation resistance
is lower than R
MAX
(see
Table2), upper to set
limit value and the
instrument generated
the Nominal test
Voltage, the instrument
emits a double sound
signal indicating the
positive outcome of
the test and displays
one screen similar to
the screen alongside.
M 05.06.02
Insulation Resistance
1.07 M
514V 60s
Voltage during the Test
Duration of the Test
Test mode
OK
TMR 500V TIME:60s
FUNZ
VNOM
Set test time
The displayed result can be stored pressing the SAVE key twice (according to
paragraph 9.1).
If the Insulation
resistance is higher
than R
MAX
(see Table2),
the instrument emits a
double sound signal at
the end of the test
indicating the positive
outcome of the test
and displays one screen
similar to the screen
alongside.
M 05.06.02
Maximum resistance value
which can be measured
(999 is displayed if a
rated voltage of 500V was
selected see Table2).
> 999 M
523V 60s
Duration of test
OK
TMR 500V TIME:60s
FUNZ
VNOM
The displayed result can be stored pressing the SAVE key twice (according to
paragraph 9.1).