6.1.2. Measurement Procedure
1. Select the desired mode by means of the F1 key.
2. Connect the black and blue test leads to B1 and B4 input terminals respectively
B2 B3
B4B1
Connection of the test leads during LOW test.
3. If the cables supplied with the instrument are not long enough for the measurement
you can extend the blue cable.
4. Short-circuit the test leads making sure that the conductive parts of the crocodiles
make a good contact to each other. Press the START key. If the display doesn't
show 0.00 repeat the test leads calibration (see paragraph 6.1.1).
5. Connect the instrument terminals to the ends of the conductor under test (see
previous picture).
6. If the mode "RT+" or "RT-" was selected use the F3, F4 keys to set the
duration of the test.
7. Press the START key. The instrument will execute the measurement. In RT+/RT-
(Timer mode) you can press START key again if you want to stop the test before the
duration set is expired.
WARNING
Never disconnect the test leads when the message "MEASURING" is
displayed.
START
STOP