TEST LEADS Before each measurement always make sure that
the calibration is referred to the cables in use.
During a continuity test, if the resistance value free
of calibration (that is the resistance value less the
calibration offset value) is negative, the symbols
is displayed. Probably the calibration resistance
value stored in the instrument memory is not
referred to the cable in use, therefore a new
calibration must be effected.
6.1.1.1. Procedure to reset test leads’ calibration parameters
To cancel calibration
parameters it is
necessary to effect a
calibration procedure
with a resistance of
test leads higher than
5 (for example with
open test leads). When
a cancellation is effected
the screen nearby is
displayed first, followed
by the screen below:
LOW 05.06.01
Message >99.9 :
means that the
instrument detected a
resistance higher
than 5 therefore it
will proceed with
Reset procedure.
>99.9
R+ R-
---- ----
---mA ---mA
AUTO 0.11
FUNC
CAL