User guide
T2000-T2100
EN – 18
4.7. LEAKAGE CURRENT MEASUREMENT (T2000)
CAUTION
Do not measure AC current values exceeding 20A in order to prevent possible
electrical shocks and any damage to the instrument.
Fig. 26: Leakage current measurement
1. Press and hold the multifunction key “1” for more than 2 seconds to switch on the
instrument.
2. The instrument displays the message “OL
”
as it automatically sets for resistance
measurement. Shortly press the multifunction
key “1” to enter current measuring mode. The
screen in Fig. 27 is displayed.
Fig. 27
3. Open the jaws and clamp the conductors corresponding to Phase and Neutral of the
single-phase system (or the earth conductor) then read the displayed result.
4.7.1. HOLD
Shortly pressing the multifunction key “3”
activates the “HOLD” function and freezes the
result on the display (see Fig. 28). To go back to
the normal measuring mode, shortly press the
multifunction key “3” again or shortly press the
multifunction key “1” (ESC function).
Fig. 28