Instruction Manual
I-V400 - SOLAR I-V
EN - 57
6.4. LIST OF DISPLAY MESSAGES
MESSAGE DESCRIPTION
Voltage too low Check the voltage between C1 and C2 input terminals
Vin > 1000 DC output voltage from module/string > 1000V
Irradiance too low Irradiance value lower than threshold limit
NTC Error Internal NTC damaged. Contact service dpt
Please wait for cooling… Instrument overheated. Wait before starting the tests again
Memory full Internal memory full. Download data to PC
Pulse width too long Anomalous condition. Repeat test with more modules
Current too low Measured current lower than the minimum detectable
Vdc wrong connection Check the voltage between C1 and C2 input terminals
Negative voltage Check the polarity of input terminals
Data base full The number of modules defined is > 30
Data @ STC unavailable The meter does not calculate the date at STC conditions
Irradiance too high Irradiance value higher than the maximum range
Data unavailable Generic error. Repeat the test
Isc too high Output current higher than maximum range
Wrong date Set a correct date/hour on meter
Error 1/2/3/4/5: contact Assistance Contact service dpt
Error EEPROM : contact Assistance Contact service dpt
Error FLASH : contact Assistance Contact service dpt
Error RTC : contact Assistance Contact service dpt
Battery low Low battery indication. Fit new batteries inside meter
Error: Vmpp >= Voc Check the settings of module inside DB
Error: Impp >= Isc Check the settings of module inside DB
Error: Vmpp * Impp >= Pmax Check the settings of module inside DB
Error: alpha too high Check the settings of module inside DB
Error: beta too high Check the settings of module inside DB
Error: Toll too high Check the settings of module inside DB
Error: gamma too high Check the settings of module inside DB
Module already present Name of module just used inside DB
Delta-Irrad. too high. Retry Unstable condition on irradiance. Repeat the test
Voltage not steady Anomalous condition. Repeat test with more modules
Unsteady Current Difference between 2 consec. Instant. values of current is > 0.13A
Firmware mismatch Problem with internal FW. Contact service dpt
Ref. Cell temp over range Temperature measured by reference cell is too high
PV module temp over range Temperature of module over maximum range
Wrong Mod. Num. Continue? Setting of number of modules not ok with measured Voc
Ref. Cell temp not detected (ENTER/ESC) Measure on the cell of module not performed
Unable to calculate Rs value Rs value external to measurement range
Thermal Instability The instrument detect unsteady conditions (see § 10.2)
Error during memory writing Problem with the access to internal memory
Error Radio: contact Assistance Contact service dpt
Error Radio transmission Contact service dpt
Error during download Contact service dpt
Recording downloaded Recording corrected downloaded on the instrument
Radio connection activated Corrected RF connection with remote unit SOLAR-02
Wait data analysis Download data from SOLAR-02 and wait for response
Unable execute analysis Problem on downloaded data from SOLAR-02. Verify settings
I < Lim Output current lower that the minimum measurable
CAUTION: internal short Contact service dpt
Remote unit not detected Enter/Esc Unit SOLAR-02 do not RF connected to the instrument
Rem. Unit in rec Unit SOLAR-02 in recording parameters Irr/Temp
STC Data available after Stop rec. Stop the recording to obtain STC data
STC Data available in MEM Data STC condition save on memory
Verify connect. P1 Check the proper insertion of test cable in the input P1
Error in data download from SOLAR-02 SOLAR I-V set for use with MPP300. Error in download from SOLAR-02.
SOLAR-02 not detected. Stop Rec.? SOLAR I-V set for use with MPP300. SOLAR-02 not detected when stopping Re
c
MPP300 not detected SOLAR I-V set for use together with MPP300. MPP300 not detected
SOLAR-02 not detected. SOLAR I-V set for use together with MPP300. SOLAR-02 not detected.
MPP300: negative AC power MPP300 has detected negative AC powers.
MPP300: inverted AC and DC voltages MPP300 has detected inverted AC and DC voltages
No power supply to MPP. Continue? Enter/Esc MPP300 has not detected the presence of the external power supply.