9. Connect irradiance probes and temperature probe (if necessary) according to previous
pictures and to effected settings (see § 6.3.2).
10. In the initial screen of IVCK mode the following values are
displayed:
Module under use
Irradiance (through direct measurement or SOLAR_02 under
RF connection)
The temperature value (if mode MAN or AUX) and
corresponding measurement mode. If mode AUTO ”- - -“
The average values of Voc and Isc under OPC conditions
The following values acquired in real time are also displayed:
Open circuit voltage
Any symbol of RF connection with SOLAR-02
15/05/10
15:34:26
Module: SUNPWR210
Irr. 980W/m2
Tc
(
AUTO
)
--
°C
VocMed
@
OPC
6
47
V
IscMed
@
OPC 5.4
3
A
Voc
@
OPC
6
4
6
V
Isc
@
OPC ---
A
Voc
@
STC ---V
Isc
@
STC ---
A
Selection
IVCK
CAUTION
When pressing the GO/STOP key the instrument can provide different error
messages (see §6.3) and, as direct consequence, not carry out the test. Check
and eliminate, if possible, causes of the problems before continuing the test.
11. Press GO/STOP to start up the test. In case of no erro
r
conditions, the instrument displays the screen with the
following results:
The module under use
The irradiance value
The cell temperature value
The average values of Voc and Isc under OPC
The values of Voc and Isc measured under OPC
The values of Voc and Isc measured under STC and relative
partial outcome resulting after comparing the rated values.
15/05/10 15:34:26
Module: SUNPWR210
Irr 932W
/
m2
Tc
(
AUTO
)
57
°C
VocMed
@
OPC
6
47V
IscMed
@
OPC 5.4
3
A
Voc
@
OPC
6
47V
Isc
@
OPC 5.
3
5
A
Voc
@
STC 7
8
7 VOK
Isc
@
STC 5.7
2
A
OK
OUTCOME: OK
Selection
IVCK
In general:


4% Isc Tol100 if
4% Voc Tol100 if
@
@@
@
@
@@
@
STC
STCSTC
STC
STC
STCSTC
STC
IscNom
IscIscNom
OKIscOutcome
VocNom
VocVocNom
OKVocOutcome
The rated values of Voc and Isc are the values of the DB built-in modules (see §5.3)
The overall value of outcome:
o OK: if all results under STC are OK,
o NO if one of the results under STC is NO
12. Press SAVE to store the test result into the instrument’s memory (see §7.2) or press
ESC/MENU to quit without saving and go back to main measurement screen.