Instruction Manual

I-V400 - SOLAR I-V
EN - 12
5.1.5. Irradiance
1. Move the cursor to “Irradiance” by means of the arrow
keys (,) and confirm with ENTER.
2. The display shows the screen with the options: “ Irr min
IV which allows to set the minimum threshold of
measured irradiance expressed in W/m
2
, used by meter as
reference during the I-V curve measurements and “Irr min
EFF” (only SOLAR I-V) which allows to set the minimum
threshold of measured irradiance expressed in W/m
2
, used
by meter as reference during the check PV efficiency
measurements
3. Set the value by means of the arrow keys (,). The
accuracy indicated in this manual is granted under
condition indicated in § 10.1 and § 10.2. In I-V curve
measurement the value is set within the 0 800 W/m
2
interval while 400 800 W/m
2
for the PV check
operations (only SOLAR I-V)
4. Confirm with SAVE and the “Saved data” message is
displayed for a while. Press ESC/MENU key to exit without
saving and to go back to the previous screen.
15/05/10 15:34:26
Irr min IV
: 0 W/m2
Irr min EFF
: 600 W/m2
SAVE to store data
SET
CAUTION
The “0 W/m2” setting for parameter “Min Irr IV” allows carrying out I-V
measurements without the following conditions being checked:
Connection of reference cell to input IRR of the instrument
Unsteady irradiation values
Number of modules consistent with the measured open circuit voltage
5.1.6. DC Clamp (only SOLAR I-V)
This option allows setting a possible correction factor K for the DC clamp, in order to improve
current measurement. If present, the correction factor is indicated on the rear label of the clamp
itself, indicated as:
K= X.xxx
In case no label is present, set k = 1.000
1. Move the cursor to “DC Clamp” by means of the arrow
keys (,) and confirm with ENTER
2. The item: “DC clamp k” is shown at display to set the
corrective factor in a interval within 0.950 and 1.050. Set
the value by means of the arrow keys (, )
3. Confirm with SAVE and the “Saved data” message is
displayed for a while. Press ESC/MENU key to exit without
saving and to go back to the previous screen
15/05/10 15:34:26
DC clamp k
: 1.000
SAVE to store data
SET