Owner manual

SIRIUS89N
EN - 93
START-UP
ANALYZER CONFIG:
All the parameters: unmodified
RECORDER CONFIG:
Start: MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key)
Stop: MANU
Integration period: 5sec
Recording of harmonics: ON
Recording of Voltage anomalies (voltage Dips and Surge) : ON
Voltage Reference for Dips and Surge detection: 230V
Upper Limit for Dips and Surge detection: 6%
Lower Limit for Dips and Surge detection: 10%
Selected voltages: V1
Selected voltage harmonics: THD, 01, 03, 05, 07, 09, 11, 13, 15
Selected currents: I1
Selected current harmonics: THD, 01, 03, 05, 07, 09, 11, 13, 15
CO-GENERATION: OFF
Selected powers, Pf and cosϕ: P1, Q1i, Q1c, S1, Pf1, dpf1
Energies: Ea1, Eri1, Erc1
POWER & ENERGY
ANALYZER CONFIG:
All the parameters: unmodified
RECORDER CONFIG:
Start: MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key)
Stop: MANU
Integration period: 15min
Recording of harmonics: OFF
Recording of Voltage anomalies (voltage Dips and Surge) : OFF
Selected voltages: V1
Selected currents: I1
CO-GENERATION: ON
Selected powers, Pf and cosϕ: P1, Q1i, Q1c, S1, Pf1, dpf1
Energies: Ea1, Eri1, Erc1
By pressing START/STOP the recording of the selected parameters is started according to
the settings made for each Typical Configuration. The rotary switch position doesn’t affect
the recording setting.