Manual
SIRIUS 87
EN - 14
4.1.3. Anomalous cases during "AUTO", "R+TIMER", "R-TIMER" tests
In case a value of Ravg or R+ or R-
higher than or equal to 5 but lower
than 99.9 was detected, at the end of
the test the instrument emits a long
sound signal and displays one screen
similar to the screen alongside.
CAL LOW
5.75
216 mA 5 s
AUTO
In AUTO mode, if a Ravg or R+ or R-
higher than 99.9 was detected, at
the end of the test the instrument emits
a long sound signal and displays the
screen alongside.
CAL LOW
> 99.9
- - - mA
AUTO
In case the mode R+TIMER or R-
TIMER was selected and a R+ or R-
higher than 99.9 was detected, the
instrument emits an intermittent
sound signal during the test, a long
sound signal at the end of the test
and displays the screen alongside.
CAL LOW
> 99.9
- - - mA 4 s
R+
TIMER
In case that:
R
MEASURED
-R
CALIBRATION
< 0 the
instrument displays the screen
alongside.
CAL LOW
0.00
219 mA
A
UTO
SAVING: The tests can be stored pressing the SAVE key twice
(according to § 5.1).
If the terminal voltage is higher than
10V, the instrument does not carry out
the test and displays the screen
alongside for 5 seconds after which, the
instrument displays the screen relative
to the precedent selected test mode
LOW under AUTO mode.
CAL LOW
13V
VOL AG
THIS RESULT CANNOT BE SAVED.
ATTENTION: value of R
avg
hi
g
her than 5
99.9 is the
maximum value
which can be
measured in the
LOW AUTO or
R+ or R-mode.
99.9 is the
maximum value.
AUTO mode.
AUTO mode.
ATTENTION: value of R
avg
or R+ or R- is too
hi
g
h.
ATTENTION:
value of R+ or R-
is too high.
Only in case the
mode R+TIMER
or R-TIMER was
selected.
Blinking CAL.
ATTENTION: R
MEASURED
-R
CALIBRATION
< 0
AUTO mode.
SAVE
ATTENTION: the
test was not
effected because
of voltage at the
terminal ends.
SAVE