Manual
MACROTEST 5035
EN - 54
If the instrument measures an
interfering voltage higher than 30V
on the amperometric circuit, it
doesn’t perform the test and a screen
similar to the one beside appears.
o 31 V
VOL AG
THE PREVIOUS RESULT CANNOT BE SAVED.
If the resistance measured is higher
than the full scale of the instrument,
when pressing START/STOP the
instrument performs the test and a
screen similar to the one beside
appears.
o >1999 Ω
- - - - - -
If the resistivity measurement is higher
than the following value
1999x6.28x(distance between the rods
selected), when pressing
START/STOP the instrument performs
the test and a screen similar to the one
beside appears.
o> 1999 Ω
- - - - - -
SAVING: The tests can be stored pressing the SAVE key twice
(according to paragraph 5.1).
Voltage value on the amperometric circuit
SAVE
1999Ω is the full
scale of the
instrument.
1999Ω is the full
scale of the
instrument.
SAVE