Manual

MACROTEST 5035
EN - 27
If the instrument detects that the earth
cable (green) is not connected, the
screen alongside is displayed for 5
seconds then the initial display is back.
Check the connections of PE conductor
under test.
RCD ~
o>1999
no PE
THE PREVIOUS RESULT CANNOT BE SAVED.
In the tests MAN x1, x2, x5 and AUTO
(during x1 and x5 tests), if the RCD
trips to separate the circuit within a time
not complying with the limits reported in
table, the instrument emits a long
sound signal at the end of the test and
displays the values alongside.
RCD ~
o 487 ms
Ι
N
30mA 1 V
x1
MAN
t
U
L
= 50V
If the RCD trips within a time higher
than the instrument’s measuring limits,
the instrument emits a long sound
signal at the end of the test and
displays the values alongside.
RCD ~
o> 999 mS
Ι
N
30mA 2 V
x1
MAN
t
U
L
= 50 V
The maximum duration depends on the test type:
Test type General RCD Selective RCD
MAN x1 test 999ms 999ms
MAN x2 test 200ms 250ms
MAN x5 test 50ms 160ms
” test 300ms
SAVING: The tests can be stored pressing the SAVE key twice
(according to paragraph 5.1).
RCD tripping time
ATTENTION: the
tripping time is
higher than the
standard limit.
RCD tripping
time:
> maximum
measuring time
(it depends on
type of test, see
following table).
Message
“no PE”: the
instrument does
not detect the
protection circuit.
SAVE
or RCD
or RCD
SAVE