Owner's manual

M71
EN – 20
6. TECHNICAL SPECIFICATIONS
6.1. TECHNICAL FEATURES
Accuracy is indicated as [% rdg + (dgt number) * resolution] at 23°C ± 5°C,< 80%HR
CAUTION
Under modes EARTH-2W and EARTH-3W the instrument displays the
symbol of warning whenever:
It is working under critical conditions, such as in the presence of input
voltages
It cannot grant the uncertainty of measurement lower than 30% of
reading, according to IEC/EN61557-1
Earth Resistance measurement
Range Resolution
A
ccurac
y
Disturbance 3Vrms
3Vrms<Disturbance<6Vrms
AUTO mode triggered
0.01 ÷ 19.99 0.01
±(2%rdg + 3dgt) ±(4%rdg + 10dgt)
20.0 ÷ 199.9 0.1
200 ÷ 1999 1
2.00 ÷ 19.99k 0.01k
20.0 ÷ 49.9k 0.1k ±(3rdg + 3dgt) ±(6%rdg + 10dgt)
Measurement frequency 110Hz ± 1Hz
Test current 12mA~
Open circuit voltage 25Vrms
Resistance values of the voltmetric (Rp) and ampermetric probes (Rc) always meet the following conditions:
¾ If R
A
10 Æ Rc, Rp 1k
¾ If 10 < R
A
500 Æ Rc, Rp 100 * R
A
¾ If R
A
> 500 Æ Rc, Rp 50k
Disturbance Voltage measurement
Range (V) Resolution (V) Accuracy
0.0 ÷ 99.9
0.1
±(2%rdg + 2dgt)
100 ÷ 299
1