Instruction Manual

HT8100
EN - 22
Resistance
Range Resolution Accuracy Output current
Protection against
overcharge
500.00 0.01 (0.2%rdg+30dgt)
1mA
1000VDC/ACrms
5.0000k 0.0001k
(0.2%rdg+10dgt)
100µA
50.000k 0.001k
10µA
500.00k 0.01k (0.5%rdg+10dgt)
1µA
5.0000M 0.0001M (1.0%rdg+10dgt)
100nA
50.00M (*) 0.01M (2.0%rdg+10dgt)
10nA
(*) Slight instability < 20 dgt
Maximum open-circuit voltage: approx 3.5V
Continuity test
Range Accuracy Buzzer Open-circuit voltage
Protection against
overcharge
500.00 (0.1%rdg+30dgt) <30
circa 3.5V
1000VDC/ACrms
Diode test
Range Accuracy
Test
current
Open-circuit voltage
Protection against
overcharge
2.000V
(1.0%rdg+10dgt)
±1mA circa ±3V 1000VDC/ACrms
AC Voltage and AC Current Frequency
Range Resolution Accuracy
Protection against overcharge
500.00Hz 0.01Hz
3dgt
1000VDC/ACrms
max 440mA
5.0000kHz 0.0001kHz
50.000kHz 0.001kHz
100.00kHz 0.01kHz
Minimum frequency value: 5Hz
Sensitivity for frequency measurements
Function Range
Sensitivity (peak to peak value)
5Hz ÷ 10kHz 10kHz ÷ 100kHz
AC mV
50.000mV
10mV 100mV
500.00mV
AC V
5.0000V 1V 1V
50.000V
1V
Unspecified
500.00V
1000.0V
AC A
50.000mA 10mA
1.000A 300mA
Generated DC Current – Programmable output
Range Resolution Accuracy
Protection against
overcharge
0.00020.000mA
0.001mA
(0.05%rdg+5dgt)
max 440mA
4.00020.000mA
Power supply: battery level > 4.5V. External power supply in simulation mode: 6V ÷ 48V
Source mode: 1.200 @20mA (battery voltage >4.5V)