Owner manual

HT4010
EN - 15
6 TECHNICAL SPECIFICATIONS
6.1 CHARACTERISTICS
Accuracy is indicated as [% rdg + dgt]. It is referred to: 23°C ± 5°C with RH <80%RH
DC Voltage
Range Resolution Accuracy
Input
impedance
Overload
protection
200.0mV 0.1mV
±(0.8%rdg + 2dgt)
10M
600VDC/ACrms
2.000V 0.001V
±(1.5%rdg + 2dgt)
20.00V 0.01V
200.0V 0.1V
600V 1V
±(2.0%rdg + 2dgt)
AC Voltage
Range Resolution
Accuracy
(50 ÷ 60Hz)
Input
impedance
Overload
protection
200.0mV 0.1mV
±(1.5%rdg + 3.5mV)
10M
600VDC/ACrms
2.000V 0.001V
±(1.8%rdg + 8dgt)
20.00V 0.01V
200.0V 0.1V
600V 1V
±(2.5%rdg + 8dgt)
AC Current
Range Resolution Accuracy
Frequency
range
Overload
protection
2.000A 0.001A
±(2.5%rdg + 10dgt)
50÷60Hz
600Arms
20.00A 0.01A
±(2.5%rdg + 4dgt)
200.0A 0.1A
600A 1A
±(4.0%rdg + 8dgt)
Resistance and Continuity test
Range Resolution Accuracy Buzzer
Overload
protection
200.0 0.1 ±(1.0%rdg + 4dgt)
150
600VDC/ACrms
2.000k 0.001k
±(1.5%rdg + 2dgt)
20.00k 0.01k
200.0k 0.1k
2.000M 0.001M ±(2.5%rdg + 3dgt)
20.00M 0.01M ±(3.5%rdg + 5dgt)
Diode test
Features Test current Open voltage
0.3mA typical 1.5VDC